NDTnet - March 1996, Vol.1 No.03
Chapter III
A particularly useful technique for measuring ultrasonic velocities in a variety of materials at elevated temperatures was established at the laboratory by Dunegan ll from a compilation of experiments performed by McSkimin,l2 and Reynolds.l3 This technique required special machining of the specimens into two concentric continuous solid cylinders of different diameters and lengths. The larger diameter cylinder was threaded with 32 threads per inch to prevent interfering reflections from the circumference. This specimen allowed simultaneous measurement of the longitudinal and shear velocities from very low temperatures to about 870°C. Special furnaces and cooling collars would have to be assembled before the test could be repeated. The technique worked out by Reynolds is shown on the following page:
Ultrasonic pulse echo waveform showing the initial pulse with longitudinal echoes and DTs.
In another technique, the ultrasonic velocities were measured through the various axes of BeO, ZnO, and CdSe, hexagonal single crystals, for the purpose of calculating the elastic constants.14 This technique was adopted from work performed by McSkimin l2 and involved the measurement of velocities through very small single crystals mounted to the end of a special quartz buffer rod. The longitudinal velocity was also measured with the aid of a water collimator-transducer system. These measurements used a tone burst generator with an adjustable pulse duration. The frequency tuning and long pulse duration resulted in ultrasonic echo interferometry. Constructive and distructive interference alternated with an increasing frequency as a result of specimen resonance.
Acoustic analysis of solid materials catagorized as thin rods, thin plates, or thin bars is often performed somewhat differently from the earlier discription on bulk materials. Sound generated into thin material may have a wavelength greater than five times the smallest material dimension. A typical test material might be wires. An analysis of the wave properties in uniform diameter thin materials such as wire is shown below:
The magnetostrictive delay line technique is used for the measurement of both extensional (compressional) and torsional (shear) wave velocities. The propagation of these waves is non-dispersive if the wavelength is greater than five times the diameter of the thin rod. The two velocities are then used, together with the mass density, to derive the various moduli. In order to receive the two interface reflections (delay line to specimen, and specimen to air) with nearly equal amplitudes for both wave velocities, a solution to a binomial equation is required. Let the delay line to specimen interface reflection be "A" amplitude, and let the specimen to air reflection be "B" amplitude and "C" the second echo from the specimen to air interface. The assumptions for the boundary conditions may be found in "Pressure Reflectivity and Transmissivity.
As a guide, the following is generally true for most metal specimens that are bonded to the magnetostrictive delay line. The following summary will be developed later in this paper. For extensional waves, the echo amplitudes from the specimen/delay line interface and from the end of the specimen, A = B when the specimen diameter is approximately half the line diameter. Then, A/B = 4 for torsional waves. Try for (A/B)ext . = (B/A)tors ., and the specimen diameter will be about 0.5 times the delay line diameter for extensional waves and about 0.7 times the delay line diameter for torsional waves. The optimum diameter ratio for extensional waves is 2:1, and the optimum diameter ratio for torsional waves is 2 1/2 :1, or 1.4:1. The average for the two calculations will be about 0.6 times the delay line diameter. Non-metal specimens may have diameters approximating the diameter of the delay line. The optimum impedance ratio is the governing factor for establishing the optimum echo amplitudes. The length of a specimen should approximate an integer multiple of a half wavelength.
This analysis is based upon an ultrasonic transducer that is attached to one end of a buffer rod and the opposite end of the buffer rod is attached to a specimen. The acoustic impedance of the buffer is Z1 ; the specimen is Z2 . The Reflection Coefficient (R) and the Transmission Coefficient (T) corresponds to the pressure wave distribution on the outgoing wave. The returning pressure waves, upon encountering the acoustic interfaces, have a Reflection Coefficient of (R ' ) and a Transmission Coefficient of (T ' ).
Let: The acoustic impedance ratio for pressure waves: r = Z2/Z1
Then: R = ( r - 1 ) ( r + 1 ); T = 2 r/( r + 1 ); T = 1 + R
R ' = ( 1 - r ) ( 1 + r ) ; T ' = 2/( 1 + r );
T ' = 1 + R ' = 1 - R
The first interface echo, A1 , returning from the buffer-specimen interface, and the subsequent echoes A2 ...AN (where N is the echo number) returning from the free end of the specimen are related in amplitude by the corresponding acoustic impedances, (Z1 , Z2) so long as the echoes are non-interfering with each other. The echo relative amplitudes, assuming that the attenuation is negligibly small, are equal to:
A1
A2 = ( 1-R² ) = ( 1-R )( 1+R ) = TT'
A3 = R'( 1-R² )
AN = R' exp ( N-2 ) TT' Where N is >= 3
The end echo of a series of repeating acoustic interfaces will be proportional to ± R' exp N TT ' , depending upon whether the termination is "open" or "short" circuited, respectively. If the attenuation is considered, the correction for the attenuation coefficient in nepers per unit length, a, is exp ( -2 aL ) A2 ; exp ( -4 aL ) A3 . Also, see Appendix C.
Where: a = ln( |R A2 A3| )/2L.
The ability to match the reflected amplitudes of the echoes returning from the buffer-specimen
interface with the echo returning from the free end of the specimen would be
desirable. Small changes in the relative diameters of the buffer or the specimen will result
in large changes in the acoustic impedance, resulting in large and unconstrained variations
in the echo amplitudes. Such amplitude variations can cause one echo to saturate and at the
same time put the other echo below the noise threshold. Thin rod, or wire, technology in
the application of ultrasound easily avails itself to adjustment of the acoustic impedance.
As long as the compressional wavelength is greater than five times the diameter of the
buffer and the specimen, the waves will propagate without dispersion. The relationship of
the diameter (D) to wavelength (
D/
Where l is the acoustic velocity divided by the acoustic frequency and the subscript
denotes the extensional (o), or the torsional (t) wave mode.
The cross sectional area or the polar moment of inertia, depending upon whether the wave
is propagating as longitudinal (extensional) or shear (torsional), respectively, influences the
acoustic impedance. For buffers and specimens that can be described as solid right circular
cylinders, the acoustic impedance is:
Zo =
Where r is the density, D is the diameter, and C is the respective sound velocity. The
impedance for torsional waves take into account the polar moment of inertia for a solid right
circular cylinder, rather than the cross sectional area used to determine the impedance for
extensional waves. Notice the strong effect the diameter has on the acoustic impedance.
The determination of the optimum reflection coefficient may proceed from the reflected
intensity relationship and the solution of the quadratic equation as follows:
Let: r = Z2 / Z1
Then: R = [ ( r - 1 ) ( r + 1 )]²
R = ( 3 ± 5 exp½ )/2 = 2.618;0. 382 (wave intensity) Solve for r in Terms of R:
r = ( 1 + R exp½ ) (1 - R exp½ )
Let:
From the assumed boundary conditions,
) may be shown as follows:
<= 0.2
D/
<= 1.2

Co D²/4 Extensional waves
Zt = 
Ct D´/32 Torsinal waves
R + T = 1
Let: A = B (equal amplitude reflections)
A = Io R (buffer/specimen interface echo amplitude)
B = Io T ² (specimen free end echo amplitude)
Io R = Io T
2
T2 = ( 1 - R )²
Io R = Io ( 1 - R ) 2 = Io ( R² - 2 R + 1 )
R = R² - 2 R + 1
Subtract R: 0 = R² - 3 R + 1
QUADRATIC SOLUTION:
R exp½ = ( r - 1 )( r + 1 ) = ±1.619; ±0.618 (wave pressure)
r = Z
: 2L/C BUFFER:= N(2thk)/C SPECIMEN , for bulk specimens also holds for thin rods with one
exception. Where N is the number of the echo; thk is the thickness of the specimen, and L
is the length of the buffer. In the application of the buffer to thin rod wave technology, the
large diameter buffer is not required. The buffer, or delay line, may be adjusted in diameter
to match the impedance of the magnetostrictive line on the one end and the specimen on the
other end. A tapered delay line is sometimes referred to as a transformer, or impedance
transformer.PRESSURE REFLECTIVITY AND TRANSMISSIVITY ²
The assumption for the following analysis is that the waves are generated in the first
medium and travel into the second medium. Some of the pressure signal is reflected at the
interface of the two mediums, but once entering the second medium, the pressure signal
propagates without reflection. Assume the boundary at X = 0, so that the exponentials
e (exp) ± jkx = 1. The total pressure is the same on both sides of the boundary; The particle
velocity into the boundary equals the particle velocity out of the boundary on the other side.
The acoustic wave pressure distribution may be developed from transmission line theory.
The subscripts 1 and 2 refer to the first and second media.
r = Z
2/Z1 (Impedance ratio at boundary)
P2+ (Pressure striking the boundary)
P2- (Pressure reflected from the boundary)
P1 = ( P1+ P1 ); P2 = ( P2+ + P2- ) (Pressure amplitudes)
uX1 = 1/Z1 ( P1+ + P1- ) ; uX2 = 1/Z2 ( P2+ + P2- ) (Particle velocity)
r P1+ - r P2 - = P2+
........ or ........ Z2 ( P1+ - P1- ) = Z1 P2+
r P1+ - r P1- = P1+ P1-
r P1+ - P1+ = r P1-+ P1-
P( r - 1 ) = P1- ( r + 1 )
P1- /P1+ = ( r - 1 ) ( r + 1 ) (normal incidence pressure reflectivity)
P2+ = P1+ + P1- (pressure partitioning)
P1- = P2+ - P1+
Z 2 ( 2 P1+ - P2+ ) = Z 1 P2+
2 Z 2 P1+ - Z 2 P2+ = Z 1 P2+
2 Z 2 P1+ = Z 2 P2+ + Z 2P2+ = P2+ ( Z 1 + Z 2 )
P2+ /P1+ = 2 Z 2/ ( Z 1 + Z 2 ) = 2 r / ( r + 1) ; (normal incidence pressure
transmissivity)
The acoustic measurements for the extensional and torsional (shear) wave velocities may be used to calculate the various moduli in homogeneous, isotropic solids. The following Table of Material Properties provides a listing of typical formulae for determining the moduli.
| Poisson's Ratio: | n = [ C²o (2 C ²t ] - 1
| Young's Modulus: | E = C²o
| Shear Modulus: | G = C²t
| Bulk Modulus: | K = C ²o C²t /[3(3C²t - C²o ) ]
| Lame Constant: | = vE /[(1 + v ) ( 1 - 2 v ) ]
Bulk Compressional Velocity: | C l = Ct | ( C ²o - 4C ²t )/(C ²o - 3C ²t )
|
| Mode | Displacement | fD = frequency X thickness, m = integer | |||||||||||||||||||||
| Symmetric | Normal | ( 2 m - 1 ) (C s / 2 ) Asymmetric | Normal | 2 m ( C s / | 2 )Symmetric | Normal | ( 2 m - 1 ) ( C l/ 2 ) | Asymmetric | Normal | 2 m ( Cl / 2 ) | Symmetric | In Plane | 2 m ( Cs / 2 ) | Asymmetric | In Plane | ( 2 m - 1 ) Cs / 2 ) | Symmetric | In Plane | mCl Cs /( C²l - C²s) exp ½ | Asymmetric | In Plane | xm Cl Cs / [ | C²l - C²s )exp½ ]* |
tan x = bx, where
and
Solve for the roots,
f; the larger the
area, the greater the coupling to fluids.
Where:
k = Radius of gyration: Plates: a /
12
a = Thickness
Y = Young's modulus
w = 2 phi f
n = Poisson's Ratio
r = density
Bars,
Where k = Radius of gyration
for rectangular bars,
where a = Thickness of bar in the direction of vibration.
k = a/2 for round bars, where a = diameter
f2 = 6. 267 f1
f3 = 17. 55 f1 For the second through fourth tone
f4 = 34.39 f1
flexural vibration mode
Plus Harmonics
= C /4L
= C /2L
Rolf Diederichs 13.Febr.1996, info@ndt.net