Effects on Refracted Angle Determination
| Table 1 Table of Modelled Probes | ||||
| Probe Diameter | Frequency | |||
| 6 | 2.25 | 5 | 7.5 | 10 |
| 9 | 2.25 | 5 | 7.5 | 10 |
| 10 | 2.25 | 5 | 7.5 | 10 |
| 12.5 | 2.25 | 5 | 7.5 | |
| 19.5 | 2.25 | 5 | 7.5 | |
| 25 | 2.25 | 5 | ||

The following set of tables summarizes the results of the graphed refracted beam angles for the modelled options. Using a Best Fit Regression Curve on the centre of beam data points the refracted angle was determined for each condition modelled. Three angles are noted in each table; the nominal angle which would be the value marked on a wedge, Snell's angle which is the angle that should result based on the velocities of the two media used, and the tabulated angles which result from the best fit curves.
| Table 2 | Table 3 | Table 4 | ||||||||||||
| Nominal 45° | Snell's Angle 44.8° | Nominal 60° | Snell's Angle 59.9° | Nominal 70° | Snell's Angle 71.2° | |||||||||
| Frequency MHz | Frequency MHz | Frequency MHz | ||||||||||||
| Probe Diameter | 2.25 | 5 | 7.5 | 10 | Probe Diameter | 2.25 | 5 | 7.5 | 10 | Probe Diameter | 2.25 | 5 | 7.5 | 10 |
| 6 | 44.1 | 44.9 | 44.7 | 45.0 | 6 | 58.5 | 59.6 | 60.2 | 60.4 | 6 | 66.4 | 69.7 | 70.7 | 71.0 |
| 9 | 44.8 | 44.7 | 44.8 | 45.1 | 9 | 59.1 | 60.0 | 60.1 | 60.2 | 9 | 68.4 | 70.6 | 70.9 | 71.1 |
| 10 | 44.5 | 45.2 | 45.1 | 44.7 | ||||||||||
| 12.5 | 44.5 | 45.2 | 45.0 | 45.0 | 12.5 | 59.9 | 60.3 | 60.2 | 60.1 | 12.5 | 69.5 | 71.1 | 71.1 | 71.2 |
| 19.5 | 45.1 | 44.3 | 45.4 | 19.5 | 60.4 | 60.2 | 59.6 | 19.5 | 70.7 | 71.1 | 71.3 | |||
| 25 | 46.2 | 46.4 | 25 | 60.2 | 59.3 | 25 | 71.1 | 71.4 | ||||||
Vertical Plane
This provides a pseudo B-scan showing relative pressure distributions in both the refracting "wedge" material and the second medium into which the beam is traveling. Media parameters used are similar to Lucite (PMMA) and steel which are commonly used in NDT applications.
1 = 1.18 g/cm3
2 = 7.8 cm3
To provide a characterisation like that obtained using manual techniques horizontal samples were taken for each probe modelled. Normally six to eight horizontal samples were used for each probe. These corresponded to depths in the second medium that are comparable to the holes found in the IOW Block used for beam profiling. At each sample depth the following values were recorded;
Since the software calculates the pressure drop for the one way transit of a beam, the 3dB and 6dB drop points on the profiles can be interpreted to represent the -6dB and -12dB boundaries normally measured in pulse-echo methods using a point reflector.
The horizontal profile of amplitude versus horizontal displacement for a particular depth is equivalent to the amplitude locus that would be traced on the CRT if a probe was moved across a side drilled hole.
|
Superimposing the -12dB DAC on the accumulated amplitude plots allows us to assess the difference between the two methods. This is shown in Figure 10.
Dropping the signal to the predrawn DAC curve results in a shift to the right for the front and back of beam data points. Although this technique is designed to correct for soundpath differences, inconsistency of results can occur due to differences in the way the DAC is drawn; e.g. number of points used, and the method in which the points are connected - straight line, best guess-hand drawn curve or electronic DAC methods. With differences of 2 to 4mm between data points the perceived accurcay advantages of using the DAC may not be as significant as intended once DAC construction variability is a considered. | ![]() Figure 10 Comparison of Two Beam Spread Determination Methods |
A probe manufacturer was given the task to manufacture an 80° refracted beam probe. The probe element was to be 2.25 MHz and 6mm (1/4") diameter. Calculations for machining the incident angle were made using Snell's Law. When the refracted angle was assessed using the standard method of a maximized peak off the 1.5mm diameter SDH in the IIW block the measured angle was around 68°. Several attempts were made to correct this by machining wedges at higher incident angles. The best that could be achieved was around 72° as determined by the peak off the 1.5mm SDH. When a 5 MHz element was used instead a slight improvement was seen but on the same series of wedges the range of measured refracted angles was 72° to 77°. [10]
A similar exercise was carried out using the RF Beam modelling for the nominal 80° refraction problem.Setup conditions for velocity and density were the same as used in the general modelling project mentioned above. Snell's Law was used to calculate several incident angles to produce refracted angles around 80° as listed below:
Modelling these angles using a 6mm diameter 2.25 MHz probe we measured the position of peak amplitude at 15mm depth just as one would do with the IIW block 1.5mm diameter hole at 15mm depth.
This was repeated with a 6mm diameter 5 MHz probe. Results are as follows;
| Table 5 2.25 MHz | Table 6 5 MHz | ||||
| Incident Angle | Snell's Refracted Angle | Max. Amp. Angle | Incident Angle | Snell's Refracted Angle | Max. Amp. Angle |
| 57° | 78.96° | 70.5° | 57° | 78.96° | 74.9° |
| 57.5° | 80.76° | 70.5° | 57.5° | 80.76° | 75.6° |
| 58° | 82.96° | 71.6° | 58° | 82.96° | 76.8° |
| 58.5° | 86.23° | 72.3° | 58.5° | 86.23° | 77.4° |
| The model accurately predicted the fact that an 80° refracted beam could never be measured using the traditional methods of determining refracted angle; no matter how high the incident angle was machined.
Figure 11 shows the beam from a 2.25 MHz 6mm diameter probe with an incident angle sufficient to produce an 86° refracted angle but the actual angle an operator would determine using the peak off a SDH at 15mm from the test surface would be 72.3°.
Figure 12 shows the beam from a 5 MHz 6mm diameter probe with an incident angle sufficient to produce an 86° refracted angle but the actual angle an operator would determine using the peak off a SDH at 15mm from the test surface would be 77.4°.
This apparent breakdown of Snell's Law is merely a result of the way in which the assessment of refracted angle is made. This is again the D/ It was concluded that when manufacturing a probe wedge it should be ascertained the range the probe is intended for and the method by which the refracted angle will be determined. This information may affect the angle the manufacturer machines the wedge angle. | ![]() Figure 11: 2.25 MHz 6mm Diameter Probe Nominal 86° Actual 72.3°
|
| Frederick H.C. Hotchkiss
is product manager for Ultrasonic Transducer Products, Nondestructive Testing Division, Panametrics Inc. in Waltham, Massachusettes, USA. He received a B.Sc. from MIT in 1966, M.Phil. and Ph.D. from Yale in 1972 and 1974 respectively. He was a postdoctoral fellow at the Smithsonian Institution in 1975. Hotchkiss is responsible for the scientific and technical development of transducers at Panametrics. Panametrics Inc. Homepage www.panametrics-ndt.com |
Ed Ginzel born 1952/04/04 in Kitchener, Ontario in Canada. Ed Ginzel is an independent consultant with the Materials Research Institute. Materials Research Institute provides training and consulting to the NDT industry. Consulting is provided to the nuclear and petrochemical industry to assist clients developing NDT techniques and to evaluate existing techniques and inspection systems. Email eginzel@mri.on.ca Homepage http://www.hookup.net/~ginzelea NDTnet Exhibition |
| | UTonline | |