
It creates RF pulse that imitates echo having fixed small amplitude and being adjustably delayed from ultrasonic instrument's initial pulse in wide time range. To use DAC Calibration Adapter the ratio between reference reflectors' echoes and coordinates of these reflectors have once to be defined using test blocks. After that DAC Calibration Adapter allows easy to create DAC curve without using any test block providing reducing deviation to value up to 1 dB only (Fig.2).
| Fig 2. Example of the DAC Calibration Adapter Operating Chart | |
| DAC curve calibration chartThis calibration chart is based on using the DAC Calibration Adapter. The DAC curve must be created using imitations of echoes from three reflectors that have amplitudes U1, U2, U3 and coordinates D1, D2, D3. U1/U2=6dB; U1/U3=11dB; U2/U3=5dB Reflectors' coordinates: D1=7mm; D2=38mm; D3=75mm * Zero of the UT instrument's digital depth meter must be previously calibrated * Sweep Delay must be adjusted equal to Zero * Sweep Range must be adjusted to 80mm | |
|
Step1. Connect DAC Calibration Adapter to the UT instrument and manipulate its delay regulator to get the 7mm UT instrument digital distance indication. Adjust amplitude of the indicated pulse to the 100% of the screen height. Mark the top of the pulse as a point 1 of the DAC curve.
Step2. Change the delay of DAC Calibration Adapter to get the 75mm UT instrument digital distance indication. Reduce the Gain of UT instrument in 11dB. Mark the top of the pulse as a point 2 of the DAC curve. Step3. Change the delay of DAC Calibration Adapter to get the 38mm UT instrument digital distance indication. Increase the Gain of UT instrument in 5dB. Mark the top of the pulse as a point 3 and create DAC curve. |
There are three human factors defining detection unreliability:[3,4]

Such presentation completely indicates operator's observing testing technology and his mistakes showing the path of ultrasonic probe acoustically coupled with the object under test. It is the first kind of NDT data. In the most cases where there is no defects in the object under test it is unique evidence of the inspection. The proposed imaging also excludes the factor of inattention to the received echoes, because the defect's projection imaging is automatically observed when the probe crosses the areas of the receiving defects' echoes. Defect's projections to the Top and Side sections of the welded joint present the second kind of NDT data: quality of object under test. To provide such presentation of the inspection results the following issues must be completed:
|
|

The main principles of I²-Sonic technology are:
|
![]() | Fig. 8 Influence of the coupling layer fluctuations on deviation DDls of the energy's transparency "probe prism to object under test" as a function of the couplant impedance z0 (Z1 - complete impedance of the probe prism for longitudinal waves; W1 - complete impedance of the probe prism for shear waves) |
It is easy to see that described I²-Sonic options provide obvious elimination operator's mistakes in the inspection procedure.

Generally, the third part of the inspection report can consist of additionally measured results plotted as much as possible automatically. To eliminatehuman mistakes in the additional measurements ultrasonic instrument must be equipped by utilities providing direct or trick easing measurement procedures. We consider that human created interpretation and conclusion must be placed in the last part of the manual ultrasonic inspection data report. Having objective automatically created presentation of calibration data, detection data, and additional measurements' data the level III specialist can interpret results obtained in the field observing, if it's necessary, postprocessing in the laboratory environment. His own conclusion about observing testing technology and quality of object under test based on the objectively obtained and presented data will grow reliability of the inspection results.