![]() ·Table of Contents ·Methods and Instrumentation | A New U-centric Radioscopy System for BGA and LSIYasushi IKEDA, Yasutoshi MIZUTAJapan Fine Ceramic Center, 2-4-1, Mutsuno, Atsuta-ku, Nagoya, 456-8587, Japan Ryusuke HIRASHIMA UNI-HITE CORPORATION, 3-28-5, Hyakunin-cho, Sinjuku-ku, Tokyo USE ELECTRONICS CO., LTD, 3-54-1, Chung Hsio East Road, Taipei, Taiwan, R.O.C. Contact |
Fig 1: Typical good joint and bad joint for BGA and contrast with various imaging methods
|
For an example, the normal radiation radiography from upward cannot detect such thin defect, which is easily understood. To obtain a detectable contrast change for such thin detachment, the most appropriate tilted viewing angle must be applied. Such appropriate viewing angles are those with large angles from the normal direction, shown in Fig.1. For a large tilted angle, the contrast difference dDb, related to the x-ray pass in the solder, Tb and the thickness of detachment, tb, is fairly larger than that for small tilted angle,dDb, which is explained by the following equation,
|
A simple method is to tilt the BGA board between the source and TV camera. But such conventional tilting is usually restricted within small limited angles between the board and X-ray generator, because a very small distance between the x-ray target and the tested board is required
for a large magnification of imaging, as shown in Fig.2(a). Whereas a CT testing can take such a large viewing angle, sometime such as 90 degree to the normal direction of a BGA board. However, the electronic board under testing is not so small, so that spatial resolution of CT becomes serious limits for the testing. This situation and the cost of the testing usually reject the use of CT testing for BGA. To realize much faster and economical method, a U-centric mechanism has been developed.
(a) Conventional tilting imaging system
|
(b) New U-centric imaging system
|
| Fig 2: X-ray imaging method for BGA | |
Figure 2(b) shows the principle of the U-centric mechanism. In this case, the BGA board is placed adjacent to the x-ray source without tilted, and an imaging TV device is moved in an x-z plane until the limit of X-ray radiation angle. The small distance between the x-ray target and the BGA can be allowed in the U-centric system for the large magnification of the testing. When the imaging device is moving, the BGA object is moved along with the x-, y- and the z-direction with keeping the constant focus point on the sample and the different magnification of imaging. Therefore, the focus point of viewing with x-rays is not changed when the viewing angle of the detector and the magnification of imaging are changed, that is, every time the viewing point is fixed on the same point of the BGA unless other viewing points are selected. This is another unique aspect of the U-centric system. With the wide movement of the TV detector, large tilting angles can be achieved, so that the small detachment at the junction of BGA can be detected. Moreover, the BGA sample can be rotated from 0 to 360 degree freely.
The commercially available U-centric 3D radiography system (EV-160m1) has been developed, in which the x-ray focal point is 1mm, and the viewing angle of the image sensor is ranging from 0 to 60 o.The achievable geometric magnification is 1000. The tube voltage is from 10 to 160 kV, and the maximum current is 0.2 mA. Radiation angle is 130o.Each solder ball of BGA can be examined without any obstruction from next balls. The system has other functions, such as image processing, psudo-3D displaying, position indicator for viewing, and measuring of dimension and area. Figure 3 shows the outlook photo of the EV-160 m1.
Fig 3: The outlook of a commercially available psudo-3D X-ray TV apparatus (EV-160m1)
|
Several examples of the BGA and other images obtained by the U-centric radioscopy system are given as followings. Figure 4 shows the x-ray images of BGA taken by the system. The BGA joints a print circuit board and LSI, only whose circuit lines can be seen by the X-ray image. The array has 400 ball joints, and each sizes of balls are about 0.8 mm. Among those ball joints, two bad joints can be seen, that is, small detachments between the pads and the balls for the joints can be seen. Figure 5 shows the U-centric images of the bad joints of the BGA taken by the same system. Two bad joints are shown with the radiation angle of 45o with various rotation angles and rotation angles. As can be seen, the detachments are clearly detected with the irradiation angle at 45oand 60o. Whereas from the normal upward direction, the detachments can not be detected, as mentioned before. In this case, the detachment of the junctions are visible clearly. From these image data, it can be concluded that the U-centric radioscopy is very effective for testing BGA joints.
Fig 4: X-ray images of BGA taken by the U-centric system
|
|
WIRE BOND
|
THROUGH HOLES
|
| Fig 5: Bad joints of BGA taken from different angles | Fig 6: Examples of X-ray image for IC circuit board taken by the system | |
Fig 7: Example of the automatic testing of a BGA board.
|
| © AIPnD , created by NDT.net | |Home| |Top| |