![]() ·Table of Contents ·Reliability and Validation | Eddy Current Probes Characterization By Means of a
G. WACHE |
If the operator disposes of only one reference tube, with artificial defects, to characterize the probes used in association with the equipement, he can be surprised by sensitivity differences observed on special events during the test, from one probe to the other.
Among other reasons depending from the probe configuration or manufacturing process , the two main listed below can explain those differences :
By using a « reference circuit », it is possible to reduce significantly the first effect and to evaluate the bandwidth of the probe associated to the equipement independantly of the reference tube : this paper shows examples of characterization results obtained for condensors and steam generators axial probes.
The probe to be tested is moved at a constant speed inside the non-conductive reference tube containing the reference circuit as shown in the Figure No 1.
Fig 1: General description of the equipement used for the tests of characterization |
Only one reference circuit is necessary for the bandwidth evaluation, but the tubes available on the market dispose of three reference circuits, as shown in the Figure No 2. We dispose of three circuits in order to compare more accurately amplitude and phase responses of the probe to be characterized, at the responses obtained with the qualified reference probe, for specific frequency values.
Fig 2: Cross section of an axial probe to be tested inside the non-conductive reference tube containing the three different « reference coils » |
The shape of the amplitude response versus frequency, corresponding to the theoretical model, is observed in the Figure No 3 showing the upper and lower limits defined for the usefull frequency bandwidth of the probe.
Fig 3: Proposal of a -6 dB criteria for usefull frequency bandwidth of the probe |
The Appendix No 2 shows the examples of experimental results obtained for different probes with different reference tubes and for two different frequency settings of the EC equipement : the Figure No 4 presents the results obtained with setting 1 for one batch of probes and the Figure No 5 presents the results obtained for another batch, in which we find the same probe reference f14,6/97-86 presented in the first batch.
Fig 4: Amplitude response of the reference coil B1 versus Frequency for four different axial probes and tubes associated with generator setting 1
|
Fig 5: Amplitude response of the reference coil B1 versus Frequency for four different axial probes and tubes associated with generator setting 2 |
The Figure No 6 gives for information, the theoretical curves obtained for three different probes with the same outside diameter but with three different coils configurations.
Fig 6: Theoretical Amplitude response of the reference coil B1 versus Frequency for three different axial probes ofconstant outside diameter |
Two main observations can be done on this Appendix No 2 :
This verification costs a lot of time, in the shop incoming process, applied for the probes ordered outside the company and we have proceeded to experiments in order to compare requirements imposed on each event of the calibration tube to the reference circuit response, for the same setting of the EC equipement, and the same scanning of the probe, as presented in the Figure No 1 of Appendix 1.
The experiment was done with two batches of qualified probes providing from two different suppliers : the Figure No 7 shows the amplitude response of the reference coil versus frequency, for two examples of those probes providing from suppliers A and B. One can observe on this figure significant differences especially for the two different frequency values reference F2 = 280 kHz and F3 = 120 kHz.
Fig 7: Amplitude response of the reference coil B1 versus frequency for two different suppliers of process |
The Figure No 8 presents the correlation observed between the amplitude response of the 7 events of the calibration tube and the amplitude response of the reference coil, for a batch of 17 probes providing from the same supplier A, and for the same gain and frequency settings of the EC equipement. One can observe for each event a linear response with a dispersion around the straight regression line generally low, excepted for the event reference VD. We can notice also on this figure the principle of the threshold defined for acceptance of the probes, tested according to this method, during the shop incoming process.
Fig 8: Amplitude response of the reference coil B2 versus amplitude response of the 7 events of the calibiration tube : selection principle |
The Appendix No4 presents the results obtained for the specific event reference VD with the 2 batches of probes tested providing from the 2 different suppliers (17 probes from supplier A and 10 from supplier B). The Figures No 9, 10 and 11 present respectively the results obtained for the 3 frequency values F1, F2 and F3. One can observe on those figures two different behaviours related to the two different suppliers, which confirm the ability of the method to separate the batches and explain some significant differences in sensitivity.
Fig 9: Amplitude response of the reference coil B1 versus amplitude response of the event VD for the frequency mode F1D
|
Fig 10: Amplitude response of the reference coil B1 versus amplitude response of the event VD for the frequency mode F2D |
In the same way, the Appendix No 5 presents the results obtained for 3 different events with the same 2 batches of probes as previously, and the same frequency setting F2. The figures Nows 12, 13 and 14 present respectively the results obtained for the 3 different event references GE 10%, 4 PHI 1 and GE LAR. We can observe that the value of RČ, reflecting the quality of the regression law, is closer to 1 for supplier B than supplier A, which confirms the ability of the method to quantify and classify the origin of the probes.
Fig 11: Amplitude response of the reference coil B1 versus amplitude response of the event VD for the frequency mode F3D
|
Fig 12: Amplitude response of the reference coil B1 versus amplitude response of the event GE 10% for the frequency mode F2D |
Fig 13: Amplitude response of the reference coil B1 versus amplitude response of the event 4PHI 1 for the frequency mode F3D
|
Fig 14: Amplitude response of the reference coil B1 versus amplitude response of the event GE LAR for the frequency mode F2D |
For the steam generator tubing, the correlation have confirmed that the amplitude response of the reference circuit was significant of the amplitude response of all the events of the calibration tube presently used and that it was possible to quantify and explain the differences observed on the two batches providing from two different suppliers of probes.
| © AIPnD , created by NDT.net | |Home| |Top| |