![]() ·Table of Contents ·Methods and Instrumentation | High Resolution Digital Flat-Panel X-Ray Detector- Performance and NDT ApplicationsJ.M. CASAGRANDE, A. KOCH, B. MUNIER, P. DE GROOTTHOMSON TUBES ELECTRONIQUES 38430 MOIRANS FRANCE Contact |
2.1 Geometrical and physical characteristics
| X | Y | Z | |
| Overall dimensions ( mm ) | 365 | 500 | 40 |
| Pixel size ( µm ) | 127 | 127 | |
| Number of pixels for the Data correction region | 2240 | 3200 | |
| Image area ( mm ) | 284.5 | 406.4 |
| Weight | 9 kg |
| Total number of pixels | 7 168 000 |
| X-ray window material | Carbon fiber |
| X-ray conversion screen | Gd2O2S : Tb |
| Conversion screen density | 34 mg/cm2 |
2.2 Electrical Characteristics
| Analog to digital converter | 12 bits |
| Image readout time | 3.2 s |
| Number of electrons per LSB | 1250 |
| RMS noise | 1.4 LSB |
| Dynamic range | >2000 : 1 |
| Non uniformity after correction | < 1 % |
| Frame time including data correction | 5 s |
2.3 Calibration and Correction
The classic procedure for image correction uses at least two calibration images : a dark image acquired without emission of X-ray photons, for offset calibration, and a so-called light image taken at the X-ray conditions close to saturation, for gain calibration.
Furthermore in order to avoid any non-linear phenomenon and to increase the correction quality, a third calibration image is acquired at X-ray conditions close to half the saturation dose.
Thanks to this bilinear correction achieved pixel per pixel, the non uniformity in the corrected image is reduced to less than 1 %.
3.1 Saturation dose
The saturation dose (ion dose in air at the entrance of the detector) varies with the spectral characteristics of the radiation. An Al filter of 2 mm thickness was added for the graph in Fig. 1.
3.2 Linearity of signal response
The linearity as a function of intensity has been measured at 60 kV and 75 kV using different thicknesses of Al absorbers. The exposure time has been kept constant. The signal response versus the absorbed dose in the phosphor screen of the detector is shown in Fig. 2. The linear range extends over a dynamic range of 103. For comparison, the signal response curve of a typical x-ray film is added which demonstrates the advantage of the flat panel detector i.e. better linearity over a wider dynamic range.
Fig 1: Saturation dose.
Parameters: Lanex Fine screen, 34 mg/cm2 of Gd2O2S/Tb, 2 mm Al filter.
|
Fig 2: Linearity for the flat panel detector and x-ray film. The X-ray film has a non-linear response and is limited in dynamic range.
|
3.3 Contrast ratio
A test pattern with different diameters of Al wires has been used (6 ISO 12 Image Quality Indicator) to determine the image contrast. Contrast is defined here as the diameter of the smallest Al wire which is still visible in the image divided by the thickness of an Al block deposited onto the test pattern. The test pattern was placed between the Al block and the detector entrance window.
The contrast ratio approaches 1% (Fig. 3), i.e. an Al wire of ¢ 0.6 mm in front of an Al block of 51 mm is still visible in the image. The signal-to-noise ratio of the visual perception is improved by the human eye because it averages along the straight lines of the wires. In general, the contrast ratio is related to the absorption ratio between wire and substrate, the scattered radiation by the substrate and the intrinsic spatial resolution of the detector. In Fig. 3 the left wing of the curve is degraded by the resolution of the detector as a result of the wire diameter close to the pixel size of 127 µm.
3.4 Modulation transfer function (MTF)
A bar test pattern of 50 µm Pb was used to obtain the contrast transfer function (CTF). The CTF values have been averaged over all possible sampling phases. From the discrete CTF values the MTF was calculated (Fig. 4). A contrast of 20% is obtained at 4 lp/mm.
Fig 3: Contrast ratio for different Al wire diameters (6 ISO 12 test pattern) at 120 kV.
|
Fig 4: Modulation transfer function (MTF) and contrast transfer function (CTF) at 60 kV, 1 mm Al filter. The Nyquist frequency given by the pixel pitch of the imager is 3.94 lp/mm.
|
4.1 NDT techniques
It is obvious that X-ray digital radiography appears to be the major technique in which the flat panel detector is involved. Due to its light weight and compact shape, this image sensor can be used in portable equipment as well as settled installation.
Nevertheless, other NDT techniques such as 3D tomography, or voludensitometry, are potential users of large area flat panel detectors, which are distortion free and not subject to magnetic fields due to moving metal parts. For instance, the acquisition time for 360 projections of 7 million pixels to obtain a tomogram should be approximately 20 minutes. However, the time for the 3D reconstruction may be much longer.
4.2 NDT fields
4.2.1 Microelectronics inspection
Fig 5: Radiography of electronic circuit.
X-ray conditions : 125 kV, 50 µA, 4 µm focal spot size. Bonding wires : 25 µm diameter.
|
4.2.2 Weld and Casting inspection
This flat panel detector is particularly well suited to the automotive industry, which is a major supplier of this domain of applications.
Detecting cracks, voids, inclusions, porosities in weld or casting parts of various sizes, such as engines or brake components, requires a good spatial resolution together with a wide dynamic range.
We may also cite the application of weld inspection on pipeline in oil industry.
The detection performances described above allow this detector to be an alternative to the radiographic film.
4.2.3 Aeronautics and space products examination
Among the various parts involved in aeronautics and space industry, the turbine blade is maybe the most inspected component, as it is a critical device of an aeronautics assembly. Images demonstrate the detection of tiny cracks in a blade, as well as the conformity control of its internal structure.
4.2.4 Composite material evaluation
The detection of delamination, internal flaw or other microstructural anomalies in complex composite materials is a well known NDT application of X-ray imaging.
Its large dynamic range associated with a good contrast transfer function allows this flat panel detector to address a large part of applications in that NDT field.
| © AIPnD , created by NDT.net | |Home| |Top| |