Krautkraemer GmbH & Co.
Robert-Bosch-Str.3
P. 0. Box 13 63
D-50330 Huerth
Phone: +49-2233-6010
Fax: +49-2233-601402
Email: hotline@krautkramer.de
Krautkramer: MAY '98 EXHIBITION

USPC 2100
Specifications


Flaw Detection Thickness Evaluation
Spike Pulser
Voltage

Energy

Risetime
Damping
Repetition Rate
Probe Type
Hi: 250 V into 500 Ohm
LO: 100 V into 500 Ohm
HI: 1470 pF
LO: 470 pF
< 10 ns
50, 75 or 500 Ohm
50 Hz - 10 kHz per channel
Single or dual element
Receiver/Amplifier
Frequency range
RF-Filters
Gain
Dynamic Range
TCG
RF Output
0.5 to 20 MHz
1, 2, 5, 10, or broad band
+80 dB
110 dB (0.1 dB steps)
35 dB, 16 segments, 12 dB/us
1 Vpp into 50 Ohm
Evaluation
Bandwidth
Digitizer

Rectification Modes
Amplitude Resolution
Time of Flight Resolution
Time of Flight Measurements


Time of Flight Modes
0.8 to 12.5 MHz (-3 dB)
50 MHz base sampling rate
200 MHz digitally up sampled
FW, HW pos, HW neg. and RF gating
1 % full screen (0 to 127 %)
0.03 mm, 0.001 inch, 0.005 µs
Inital pulse to gate A and B
Interface to gate A and B
- -
Flank or peak


800 MHz digitally up sampled
RF gating
- -
0.004 mm, 0.00015 inch, 0.00125 µs


Gate A to gate B

Outputs
Analog Amplitude
Analog Time of Flight

Alarms
Output Update Rate
2, 5 or 10 V full scale (8 bit)
2, 5 or 10 V full scale (12 bit)
Selectable offset and range
TTL active Hi or active Lo
Rep rate, timed, or latched alarms
no




A-Scan Display
Modes
Gates
TCG Curve
Range


Delay


Trigger Modes
Frozen A-Scan
Traces
FW, HW pos, HW neg. and RF
Bar display with delay, width and threshold
Displayed midscreen or baseline
3 to 1000 mm
0.05 to 50 inch
1 to 321 us
-25 to 975 mm
-1 to 49 inch
-10 to 310 us
Inital pulse, artificial, or interface echo
Superimposed on live A-Scan
1, 2, 4 or all
Gates
Number
Thresholds
Limits
Noise Suppression
Synchronization Modes
Measurements



Range


Delay


1 interface and 2 measurement
1 per gate (selectable polarity)
- -
Selectable violations per window
Initial pulse, artificial, or interface echo
Current amplitude
Minimum/maximum/average amplitudes
Flaw depth
Threshold evaluation (alarms)
1 to 1000 mm
0.05 to 50 inch
1 to 321 us
0 to 1000 mm
0 to 50 inch
0 to 320 us

- -
Minimum wall and maximum wall


Current thickness
Minimum/maximum/average thickness
- -
- -







NDTnet 1998 May, Vol.3 No.5
SPECIAL MAY '98 EXHIBITION
Ultrasonic Flaw Detectors
| NDTnet Exhibition |
|Index of Exhibitors| |Index of Showcases|
| NDTnet |
1998 ©copyright Rolf Diederichs, info@ndt.net
/DB:Exh /SO:NDTnet /IN:Krautkramer /CN:DE /CT:UT /CT:instrument /CT:Flaw Detector /ED:1998-5